Plastic
TTM-MI-TP Melt Flow Indexer
This machine is used to determine the MFI value of the melt flow rate through a standa...
Detail
TTM-MI-TP Melt Flow Indexer

Features
This machine is used to determine the MFI value of the melt flow rate through a standard die every 10 minutes under a viscous flow state and a certain temperature and load. It is suitable for polycarbonate, polyarylsulfone and fluorine with higher melting point. Engineering plastics, such as plastic and nylon. It is also suitable for the detection of low melting temperature plastics such as polyethylene (PE), polystyrene (PS), polypropylene (PP), ABS resin, polyoxymethylene (POM), polycarbonate (PC) resin, etc. It is widely used in plastic production, Plastic products, petrochemical industry and related universities, research institutes and commodity inspection departments.

Testing Standards
  •  GB/T 3682-2018 "Determination of Thermoplastic Melt Flow Rate and Melt Volume Flow Rate"
  •  ISO 1133: 1997 "Determination of Thermoplastic Melt Mass Flow Rate MFR and Melt Volume Flow Rate MVR"
  •  ASTM D1238 "Standard Test Method for Measuring Thermoplastic Melt Flow Rate by Extrusion Plastometer"
  •  DIN 53435 "MFR measurement method of melt index instrument"
Technical Parameters
üTest method: quality method
üDisplay mode: touch panel controller
üColor LCD display only in Chinese and English
üAutomatic manual cutting
üDual temperature control system, the temperature control accuracy is more accurate, and the heating efficiency is higher
üDouble imported sensors, the upper and lower temperature gradients are more accurate, can be used independently, and improve the service life
üThe barrel is made of imported Hastelloy alloy nitriding, and the die is made of carbon tungsten steel. The material used is real
üThe heat preservation barrel is made of all sus304 stainless steel, which will not rust at high temperature for a long time
üAutomatically calculate the result and print it out
 
Main Technical Specifications
lTemperature range: 40℃-450℃
lTemperature drift: 4 hours≤0.5℃
lTemperature fluctuation: ±0.2℃
lTemperature uniformity of 10 mm-70 mm above the die: ±1℃
lTemperature display resolution: 0.1℃
lTime display resolution: 0.1 S
lCutting timing range: 1-999 S adjustable
lDie diameter: Φ2.095±0.005 mm (half die 1.05 mm+0.005, optional)
lLength of discharge port: 8.000±0.025 mm
lPiston rod diameter: 9.474±0.007 mm
lPiston rod length: 6.35±0.13 mm
lDiameter of charging cylinder: Φ9.550±0.007 mm
lTest range: 0.1-150 g/10 min
lWeight accuracy: ±0.5%
lOutput mode: micro automatic printout
lCutting method: manual and automatic cutting
lTest load: a total of 8 levels, 8 sets of weights
lPower supply voltage: AC 220 V±10%; 50 Hz
 
List of Tools
No. Name Unit Quantity Remarks
1 Frame set 1 Including printer
2 Load set 8 grades, including 0.325 kg1.200 kg2.160 kg3.800 kg5.000 kg10.000 kg12.000 kg21.6 kg (1# within loading class 1)
3 Load tray piece 1 within loading class 1
4 Piston rod piece 1 within loading class 1
5 Die piece 1 Tungsten carbide
6 Loading hopper piece 1 Stainless steel
7 Loading rod piece 1  
8 Barrel cleaning rod piece 1  
9 Die cleaning rod piece 1 Stainless steel
10 Gauze roll 2  
11 Printer paper roll 2  
12 others Operational manual in English x 1 copy,  feeding tray x 1