TTM-VIA series Auto Vickers Hardness Tester
Product Briefings
Automatic Vickers hardness tester, load range 980.7 mN-490.3 N electronic closed-loop control Optional test method Automatic turret Can be equipped with 4 objective lenses Auto focus, automatic reading LED indicator Small and light Windows operating system
Features
Flexible autofocus, automatic reading,
lFour sets of complex test piece surfaces can be pre-set for automatic focusing and automatic reading memory settings. If the same complex surface is encountered, the automatic focusing and automatic reading can be accurately performed as long as the memory setting is executed.
lIt is very difficult to perform auto-focusing and auto-reading on surfaces that are not smooth and full of scratches, or through corroded surfaces. This function allows the system to pre-set the focus and reading capabilities on difficult surfaces through pre-prepared test strips. Improve the accuracy of focusing and reading, increase the test speed
Specifications
Model | VIA-F | VIA-S | ||
Main loading | Vickers | 0.981, 1.961, 2.942, 4.903, 9.807, 19.61, 29.42, 49.03, 98.07, 196.1, 294.2, 490.3 (N) 0.1, 0.2, 0.3, 0.5, 1, 2, 3, 5, 0, 20, 30, 50 (Kgf) |
||
Knoop | 0.981, 1.961, 2.942, 4.903, 9.807, 19.614 (N) 0.1, 0.2, 0.3, 0.5, 1, 2, (Kgf) |
|||
Brinell | 147.1, 294.2,441.3 N (15,30,45 Kgf) | |||
Rockwell | 588.4, 980.7, 1471 N (60,100,150 Kgf) | |||
Brinell | 1 / 9.807, 1 / 24.52, 1 / 98.07, 1 / 294.2, 2.5 / 61.29, 2.5 / 153.2, 2.5 / 306.5, 5 / 245.2 N 1/1, 1 / 2.5, 1/5, 1/10, 1/30, 2.5 / 6.25, 2.5 / 15.625, 2.5 / 31.25, 5/25 (Kgf) |
|||
Loading speed | 0.981-4.903 N; 50 um / 9.807-490.3 N; 100 um/s | |||
Indenter | Standard; Vickers (HV), optional Knoop (HK), optional Brinell (HBW1, 2.2, 5 mm) | |||
Objective lens | Standard 10X, 20X optional 5X, 50X, 100X | |||
Test methods | Fully automatic / manual autofocus | |||
Min. test unit | 0.1nm | |||
Auto focus speed | 2 seconds / point (related to samle surface) | |||
Camera | Standard 1.3-metapixel USB camera | |||
Work bench | 50x50 mm XY automatic table, resolution 1 um | |||
Control box | Three axis control box | Single axis control box | ||
Operation panel | All-in-one PC | Desktop PC | ||
Max. sample height | 140 mm | |||
Max. sample depth | 100 mm | |||
Dimensions | W200xD370xH500 mm | |||
Weight | 34 kg | |||
Standards | ISO6507-2,JIS B7725,JISB7734,ASTM E92,ASTM E384 | |||
Power | AC100-240V 50/60HZ |